Difference between revisions of "Compartment pressure measurement"
Ostermayer (talk | contribs) |
Ostermayer (talk | contribs) (→Source) |
||
Line 25: | Line 25: | ||
==External Links== | ==External Links== | ||
− | == | + | ==Sources== |
− | + | *Image and procedure text cortesy of SoBro EM http://sobroem.com/2014/04/09/under-pressure/ | |
<references/> | <references/> | ||
[[Category:Procedures]][[Category:Ortho]][[Category:Trauma]] | [[Category:Procedures]][[Category:Ortho]][[Category:Trauma]] |
Revision as of 19:27, 20 July 2014
Contents
Indications
Contraindications
Equipment Needed
Procedure
- Contents of the pressure measurement kit
- Remove contents from wrapping
- Attach chamber to the pre-filled saline syringe
- Place the aforementioned into the monitor/unit
- Place needle onto chamber
- Eject excess air, if present, from syringe
- Zero the assembled unit at the angle you will be entering compartment
- Insert, inject 2-3 drops of saline, and await measurement
Pressure Interpretation
Interpretation of Compartment Pressure
- Normal is <10 mm Hg
- Pressures <20 mmHg can be tolerated w/o significant damage
- Exact level of pressure elevation that causes cell death is unclear. It was previously thought pressure >30 mmHg was toxic although the "delta pressure" may be better predictor than absolute pressure
ΔPressure = [Diastolic Pressure] – [Compartment Pressure][1]
- ΔPressure < 30 mm Hg is suggestive of compartment syndrome
Complications
See Also
External Links
Sources
- Image and procedure text cortesy of SoBro EM http://sobroem.com/2014/04/09/under-pressure/