Ocular ultrasound
Technique
- Use vascular/linear probe
- Plenty of ultrasound gel to decrease amount of pressure needed to place on eye
Elevated ICP
- Measure optic nerve 3mm posterior to the globe, from inner wall to inner wall
- Normal is <5mm
Globe Rupture
- Only perform if you can ensure that you do not put pressure on the globe
- Findings
- Decrease in size of globe
- Anterior chamber collapse
- Vitreous hemorrhage
- Buckling of the sclera
- see Globe Rupture
Intraocular Foreign Body
- Bright, echogenic acoustic profile w/ associated shadowing or reverberation
Retinal Detachment
- Echogenic undulating membrane in the posterior globe, protruding into the vitreous
- Evaluate with patient moving eye left/right
- SN 97-100% and SP 83-100%[1]
Vitreous Hemorrhage
- Vitreous filled with multiple large echoes
- Increasing the gain is helpful for detecting acute hemorrhages