Transvenous pacing
Indications
- Faiure of transcutaneous pacing + chronotropes
- Sinus Arrest/Bradycardia
- AV Block - 3rd or 2nd degree. May occur from MI, bacterial endocarditis, or Lyme dz
- Drug OD - Digoxin toxicity
- Overdrive pacing
Contraindications
- Asystolic cardiac arrest
- Hypothermia bradydysrhythmias
Procedure
- Site selection: Right IJ or left subclavian for most smooth anatomical course
- Equipment:
- External generator: Shows rate (bpm), output (mA) and sensitivity
- Cordis Kit (7 Fr): Gold strip on kit at Harbor. Contains cordis introducer, TV pacer catheter and pacer wire sheath.
- Similar to a central line, place single lumen catheter under ultrasound.
- Pacing catheter
- Test small balloon for leaks prior to insertion with 1.5mL of air while balloon rests in a container of saline
- Connect the positive and negative electrodes to the external generator
- Advance the catheter through the introducer sheath to roughly 20 cm and insert through cordis diaphragm
- External generator: Set HR 80, start at max current output (usually 20 mA), and sensitivity all the way down (paces no matter intrinsic rate)
- Pearl - Digital generators have "emergency" button that goes to necessary settings
- Advancing Pacer Wire:
- Blind approach: Inflate balloon and advance slowly until you see pacer spikes on monitor followed by a widened QRS (similar appearance to LBBB).
- Ultrasound approach: Have assistant give a subcostal/parasternal long axis, which gives visualization image of when electrode contacts the wall of the RV.
- Sensing approach: Use alligator clip to connect negative pacer electrode to any precordial lead. Look for ST elevation when RV endocardium engaged.
- Fluoroscopy: If time permits, use this method in a patient that has prior pacemaker/AICD. Placement of TV pacer w/o fluoro can disrupt prior electrode placements.
- Deflate balloon (leave syringe attached) and secure catheter in place
- Make sure swan sheath fully extended, and locked onto cordis hub
- Take note of pacer depth in case it's accidentally moved
- Final Settings
- Current: Determine threshold level by reducing electrical current settings until capture lost. Final current set to twice the threshold level for pt
- Sensitivity: Adjust so only paces when necessary (not too high or too low) to allow intrinsic beats and supplement if needed. Do not want oversensing or undersensing.
- Placement confirmation
- Good to obtain baseline CXR, which should show the catheter tip over the inferior border of the cardiac shadow
- EKG shows paced QRS exhibiting a LBBB morphology, and a superior QRS axis
Complications
- Related to central venous access
- Infection, PTX, air embolism, arterial puncture and venous thrombosis
- Related to pacing catheter:
- Valvular tear(s)/rupture(s)
- Myocardial Peforation (atria/ventricle/septum) - consider tamponade
- Ventricular Arrhythmias: VT or VF
Links
- University of Cincinnati
- Drexel Video
- Practical Pointers